Luosh Mask Testing Data

Select Your Mask Carefully - Only Use ASTM-Rated Masks That Meet Proven Standards

When safety is your concern, choosing the right mask is more than a matter of preference. You have a lot at stake when selecting proper protection for your application.  Know how you plan to use your facial protection, and select an ASTM-rated mask that is right for you and offers the best protection.

Luosh American Made Face Masks are rated:

ASTM F2100-11 Level III

ASTM International - Leading The Way

ASTM International sets international standards for healthcare best practices, which includes testing and requirements for performance of materials.


The FDA, who oversees marketing and sales of all surgical masks, recommends that 4 standards of mask performance be met:

  1. Fluid Resistance
  2. Differential Pressure (Delta-P)
  3. Filtration Efficiency
  4. Flammability

 

Luosh Face Mask Testing Data

Nelson Labs Study No. 1361719-S01  

Fluid Resistance: Pass at 160 mmHg (21.3 kPa) - Download Report
  • Pre-Conditioning: Minimum of 4 hours at 21 ± 5°C and 85 ± 5% relative humidity (RH)
  • Test Conditions: 23.8°C and 21% RH
Differential Pressure: 4.9 Delta P mm H2O/cm2 | 48.64 Pa/cm2 - Download Report
  • Test Side: Inside
  • Delta P Flow Rate: 8 Liters per minute (L/min)
  • Conditioning Parameters: 85 ± 5% relative humidity (RH) and 21 ± 5°C for a minimum of 4 hours
Particle Filter Efficiency: 99.75% - Download Report
  • Test Side: Inside
  • Area Tested: 91.5 cm2
  • Particle Size: 0.1 μm
  • Laboratory Conditions: 21.3°C, 22% relative humidity (RH) at 2128; 21.3°C, 22% RH at 2221
  • Average Filtration Efficiency: 99.75%
  • Standard Deviation: 0.060
Bacterial Filter Efficiency: 99.99% - Download Report
    • The challenge delivery was maintained at 1.7 - 3.0 x 103 colony forming units (CFU) with a mean particle size (MPS) of 3.0 ± 0.3 μm.
    • Area Tested: ~40 cm2
    • BFE Flow Rate: 28.3 Liters per minute (L/min)
    • Laboratory Conditions: 85 ± 5% relative humidity (RH) and 21 ± 5°C for a minimum of 4 hours
    • Test Article Dimensions: ~175 mm x ~157 mm
    • Positive Control Average: 2.7 x 103 CFU
    • Negative Monitor Count: <1 CFU
    • MPS: 2.9 μm
    Flammability: Class 1 - Download Report
    • Test article ignited, but extinguished.  Exceeds standard requirements.